OPERATION AND MEASUREMENT

Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying
or recording a voltage signal with a range of 0-2.5 V (SQ-202X) or 0-5 V (SQ-205X) to cover the entire range of PPFD
from the sun. In order to maximize the measurement resolution and signal-to-noise ratio, the signal input range of
the measurement device should closely match the output range of the quantum sensors. The amplification circuit
requires a power supply of 2.5-24 VDC (SQ-202X) or 5-24 VDC (SQ-205X). NOTE: To prevent sensor damage,
NOT connect the sensor to a power source greater than 24 VDC.
Wiring for SQ-202X and SQ-205X
White: Positive (signal from sensor)
Red: Input Power SQ-212X 3.3-24 V DC, SQ-215X 5.5-24 V DC
Black: Ground (from sensor signal and output power)
Clear: Shield/Ground
DO