Cosine Response:
Directional, or cosine, response is defined
as the measurement error at a specific
angle of radiation incidence. Error for
Apogee SQ series quantum sensors is
approximately ± 2 % and ± 5 % at solar
zenith angles of 45° and 75°, respectively.
Mean cosine response of twenty-three SQ
series quantum sensors (
represent two standard deviations above
and below mean
measurements were made by direct side-
by-side comparison to the mean of four
reference thermopile pyranometers, with
solar zenith angle-dependent factors
applied to convert total shortwave
radiation to PPF. Blue points represent the
AM response and red points represent the
PM response.
error bars
). Cosine response
8