SS-110 Sensitivity
SS-120 Sensitivity
Sensitivity was determined by collecting
spectra under a quartz halogen lamp with an
NIST traceable calibration and calculating the
ratio of counts measured with the
spectrometer to energy flux density from the
lamp. Maximum sensitivity of the SS-110 is
near 600 nm. Sensitivity of the SS-110 is at
least 10 % of the maximum value at all
wavelengths greater than 380 nm.
Sensitivity was determined by collecting
spectra under a quartz halogen lamp with an
NIST traceable calibration and calculating the
ratio of counts measured with the
spectrometer to energy flux density from the
lamp. Maximum sensitivity of the SS-120 is
near 750 nm. Sensitivity of the SS-120 is at
least 10 % of the maximum value at all
wavelengths less than 1030 nm.