Apogee Instruments SU-300 UV-I Owner's Manual - Page 12
Browse online or download pdf Owner's Manual for Accessories Apogee Instruments SU-300 UV-I. Apogee Instruments SU-300 UV-I 19 pages. Ultraviolet index & ultraviolet-b sensor
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying
or recording a millivolt signal. An input measurement range of approximately 0-2 mV is required to cover the
entire range of the SU-300 UV-I/UV-B sensor. To maximize measurement resolution and signal-to-noise ratio, the
input range of the measurement device should closely match the output range of the UV sensor. DO NOT connect
the sensor to a power source. The sensor is self-powered and applying voltage will damage the sensor.
Wiring for SU-300
White: Output Signal
Black: Ground
Clear: Shield