DataApex Shimadzu LC-20 Manual - Page 14

Browse online or download pdf Manual for Water Pump DataApex Shimadzu LC-20. DataApex Shimadzu LC-20 19 pages. Metal free pump

4 Using the control module

4.1.1 Gradient Options

Invoke the Options... button in the
open the
name for particular solvents, to switch whether they are used or not in the
gradient and to set the warning levels for pressure to prevent the damage
to hardware.
Min. Pressure
Sets the minimum pressure for the given pump. When pressure drops to
the set value, the pump will shut down. This prevents the solvent leakage.
Max. Pressure
Sets the maximum pressure for the given pump. When pressure reaches
the set value, the pump will shut down. This serves to prevent the damage
to the pump when the column is blocked.
Solvent 1 (..4)
It is possible to enable/disable particular solvent, as well as to set custom
name to it.
Gradient Options
Fig 6: Gradient Options
Method Setup - LC Gradient
dialog. This dialog allows to set the custom
- 10 -
Clarity Controls
dialog to