Presetting parameter

Paramete
Code
r
P1.xx
00-15
P2.xx
0/1/2/
3-50
P3.xx
0/1...5
P4.xx
0/1/2/
3/4
P5.xx
0/1/2
P6.xx
00-31
P7.xx
0/1/2
P8.xx
0/1/2/
3/4/5
P9.xx
0/1
P10.x
0/1/2/3/
4/5/6
P11.x
0/1/2/3/
4/5/6/7
11
Presetting
00: no auto-off
01-15: minutes auto-off time
0: no hold function
1: hold larger weight reading
2: auto release hold function when weight is below 10d
and auto-hold new stable weight (more than 10d) (nur
EHR4)
3...50: unchangeable reading when the variety is within +/-
3d-50d
0: no RS232 function
1: continuously output display data
2: continuously output gross, net and tare weight
3: output display data one time when scale is stable
4: output gross, net and tare weight one time, when scale
become stable
5: Bio-RS232
Baud-Rate for RS232 interface
(Modifications in setting only by authorized personsl!)
0: 1200bps, 1: 2400bps, 2: 4800bps, 3: 9600bps, 4:
19200bps
RS232 interface
(Modifications in setting only by authorized personsl!)
0: 8N1, 1: 7O1, 2: 7E1
Resolution
(Modifications in setting only by authorized personsl!)
500, 600, 750, 800, 1000, 1200, 1500, 2000, 2400, 2500,
3000, 3500, 4000, 5000, 6000, 7000, 7500, 8000, 10000,
12000, 15000, 20000, 25000, 30000, 35000, 40000, 50000,
60000, 70000, 75000, 80000, 100000
Division select
0: 1, 1: 2, 2: 5
Decimal point in calibration
(Modifications in setting only by authorized personsl!)
0: x1, 1: x0.1, 2: x0.01, 3: x0.001, 4: x0.0001, 5: x10
Calibration unit
(Modifications in setting only by authorized personsl!)
0: kg, 1: lb
Weiging units enabled
0: only kg, 1: only lb, 2: only lb:oz, 3: kg or lb, 4: kg or lb:oz,
5: lb or lb:oz, 6: kg, lb or lb:oz
Power-on zero-point range
(Modifications in setting only by authorized personsl!)
Calibration zero-point 0: +/-1%FS, 1: +/-2%FS, 2: +/-5%FS,
3: +/-10%FS, 4: +/-20%FS, 5: +/-50%FS, 6: +/-100%FS,
7: no limitation
EHR_EHQ-210517-Rev009-UM-en