Conviron CMP6010 Quick Start - Page 2

Browse online or download pdf Quick Start for Controller Conviron CMP6010. Conviron CMP6010 2 pages.

Humidity
Temperature
Light Level
PROGRAM: a Program is a collection of lines that specify chamber
environmental conditions in real time.
To enter a program:
1. Navigate to the PROGRAM menu and select CREATE
2. Choose a number under which to save the program (controller
default = lowest available number)
3. Press < > with the cursor at timeline 01 and begin entering set
point values until the program is complete
4. Press <Esc> to save the program
SCHEDULE: the Schedule is a series of entries, each of which specifies
which program to run and how many times to repeat that program.
To create a schedule:
5. Navigate to the SCHEDULE menu and press EDIT
6. Press < > to select entry 01, and choose the program you wish to
run, and the number of times you would like that 24 hour program
repeated. One repetition will run the program one time.
7. Press <Esc> and choose the number of times you would like the
whole schedule to be repeated
To run the schedule:
8. From the SCHEDULE menu select RUN
9. Using the <
/
> keys, choose the entry you wish to be the
starting point of the schedule (only one in example program)
10. If you have chosen more than one repeat, choose from which
repetition you would like to begin the schedule
Copyright © 2013 by Controlled Environments Limited. All rights Reserved.
01-28-13 Rev. 00
EXAMPLE PROGRAM
01
02
30
25
20
15
10
5
0
-5
00:00
4:00
Program Timelines
03
04
8:00
12:00
16:00
Time [hh:mm]
Line
{1 00 : 00 08.0 045
02 05 :00 15.0 055
03 09 :30 25.0 060
0 4 14 :00 22.0 055
<
Line
{5
06
07
08
<
01
02
<
05 06
20:00
00:00
E D I T P R O G R A M # : 0 1
> to Edit Line
Example Program: Screen 1
E D I T P R O G R A M # : 0 1
18: 00 15.0 050
20: 00 10.0 045
> to Edit Line
Example Program: Screen 2
E D I T S C H E D U L E
1
7
> to Edit Entry
Example Program Schedule:
Program 1 set for 7 repetitions
2 of 2
60
50
40
30
20
10
0
-10
0
1
3
2
1
0