ADE M501020 Operating Manual - Page 14

Browse online or download pdf Operating Manual for Scales ADE M501020. ADE M501020 20 pages. Electronic wheelchair scales

ADE M501020 Operating Manual

Cleaning

Please disconnect the unit from the operating voltage before cleaning. Do not use any aggressive
cleaning materials for cleaning; only use a damp cloth or conventional disinfectant.
Make sure that no liquids penetrate the unit!

Error messages

Error mes-
Description
sage
Overload display
Overflow indicator (negative)
Overflow indicator (positive)
AD LOW
A/D above negative range
AD HI
A/D above positive range
EE SUM
Parameter or calibration error
EE WR
EEPROM indicates an error
PM SUM
Internal program indicates a check error
HOFSET
Switch on – above zero
LOFSET
Switch on – below zero
UOFSET
Irregular

Safety instructions

Hazard-free operation is no longer guaranteed
-
if the power unit exhibits visible damage.
-
if the power unit is no longer working.
-
after longer periods of storage in damp rooms.
Inform your service partner in such cases for your own safety.
Any power units or batteries other than those supplied by the manufacturer may not be
used. There is otherwise a risk of influencing other medical, electrical devic-
es/equipment!
If the chair scales remain unused or the patient must wait while remaining seated in the
chair scales, secure the unit by operating both brake locking devices on the rear
wheels.
The connecting cable between the scales and power unit must be laid in such a way
that they do not present a trip hazard.
14

Care, servicing and recalibration

Problem solution
Weight exceeds max. capacity of the
scales.
Negative weight indicated – weight too
low to be displayed (<=99999).
Positive weight indicated – weight too
high to be displayed (>=999999).
Check if the scales are correctly as-
sembled or if they are damaged.
Check if the scales are correctly as-
sembled or if they are damaged.
Recalibrate – please contact your sup-
plier or service partner:
Please contact your supplier or service
partner.
Please contact your supplier or service
partner.
Please remove the weight from the
scales and restart.
Check scales for damage and inform
your service partner:
Do not touch the scales after they have
been switched on.
M501020-210322-Rev006-UM-e