dymax 39560 User Manual - Page 8
Browse online or download pdf User Manual for Measuring Instruments dymax 39560. dymax 39560 14 pages.
Figure 1. Spot Radiometer (PN39560)
Operation
1. The ACCU-CAL™ 50 Radiometer may be used to measure UV intensity from Flood Lamps or UV
intensity from the end of a 3-mm, 5-mm, or 8-mm lightguide used with a Spot Lamp. For Flood Lamp
use, attach the Detector to the Radiometer as shown in Figure 3.
2. For Spot Lamp use, select an Adapter that matches the size of the Lightguide that is installed on the
Spot Lamp. 3-mm, 5-mm, and 8-mm Adapters are available.
NOTE:
Lightguide Adapters are included in the spot version (PN 39560) of the ACCU-CAL™ 50 or
available separately.
Figure 3. Radiometer with Detector
3. Install the Lightguide Adapter on the end of the Detector using the two, 2 mm screws provided.
Dymax ACCU-CAL™ 50 Radiometer User Guide
Figure 2. Flood Radiometer (PN39561)
Figure 4. Lightguide Adapters
7