dymax ACCU-CAL 150 User Manual - Page 6
Browse online or download pdf User Manual for Measuring Instruments dymax ACCU-CAL 150. dymax ACCU-CAL 150 20 pages. Uv radiometer
6
Dymax ACCU-CAL™ 150 Radiometer User Guide
Product Overview
Description of the ACCU-CAL™ 150 Radiometer
The Dymax ACCU-CAL™ 150 UV Radiometer measures and displays the total UV energy and UV irradiance of a
UV light-curing system. Its compact size and robust design withstands the extremes of UV light-curing
environments while providing accurate measurement.
The carefully designed optical sensing system measures the wavelengths that are relevant to the UV process.
The output of the sensing system is converted to digital form and displayed on an easy-to-read LED display.
The ACCU-CAL™ 150 measures ultraviolet wavelengths defined as UVA (320-390 nm), and displays those
readings as total energy and dosage.
Environmental Considerations
Optical measurement instruments are sensitive to extremes in environmental conditions like high
temperature, humidity, and contamination. Protect the device and its detector(s) from high humidity,
high temperature, direct sunlight, and contamination.
Do not use the ACCU-CAL™ 150 immediately after moving it from a cold to a warm environment. Under
certain circumstances, condensation could develop that may cause inaccurate measurement results.
Allow the device to adjust to room temperature before use.
Do not use the ACCU-CAL™ 150 in powerful magnetic, electromagnetic, and electrostatic fields. These
disturbances may influence measurement results.
Do not expose the ACCU-CAL™ 150 to temperatures exceeding 65°C during an exposure run. Exposure to
temperatures exceeding 80°C will cause unit failure and will void the warranty.