dymax BlueWave MX-150 User Manual - Page 6

Browse online or download pdf User Manual for Laboratory Equipment dymax BlueWave MX-150. dymax BlueWave MX-150 16 pages. Ledd spot-curing emitters
Also for dymax BlueWave MX-150: User Manual (12 pages)

6
BlueWave® MX-150 Emitter User Guide

Unpacking

Upon arrival, inspect all boxes for damage and notify the shipper of box damage immediately.
Open each box and check for equipment damage. If parts are damaged, notify the shipper and
submit a claim for the damaged parts. Contact Dymax so that new parts can be shipped to you
immediately.
WARNING!
via the interconnect cable it is susceptible to ESD damage, handle according to
ESD standards using a ground strap and do not touch exposed connector pins.
The parts below are included in every package/order. If parts are missing from your order,
contact your local Dymax representative or Dymax Customer Support to resolve the problem.

Parts Included

LED Emitter
BlueWave® MX-150 LED Emitter Assembly
5-mm Lightguide Simulator
User Guide

Installation

The BlueWave® MX-150 emitter is part of a MX-series curing system and requires connection
to a controller via an interconnect cable for proper operation.

Important Information

Do not connect any components while power is applied.
Mount the BlueWave® MX-150 emitter to a rigid support, such as the emitter
stand PN 42390, prior to connecting the interconnect cable to prevent handling
damage.
Do not touch the emitter aperture glass. This can result in poor performance
and broken glass due to heating. Inspect before each use and clean with
isopropyl alcohol if contaminated.
Until the BlueWave® MX-150 emitter is attached to a controller