Active Motif PIXUL User Manual - Page 4

Browse online or download pdf User Manual for Medical Equipment Active Motif PIXUL. Active Motif PIXUL 15 pages. Multi-sample sonicator

General Warnings

Don't run the PIXUL instrument without PIXUL Coupling Fluid.
Tilting the PIXUL instrument can result in spillage of Coupling Fluid.
Do not unplug or power off the PIXUL instrument during a run. Loss of power to the instrument
while a file is open can result in corruption of the data file and operating system.
Change PIXUL Coupling Fluid every 2 months.
PIXUL Coupling Fluid temperature should be approximately 15°C at the start of the sonication run.
Don't exceed a Coupling Fluid temperature of 35°C during sonication. If Coupling Fluid tempera-
ture is maintained within this range, in-well thermal monitoring indicates that the temperature of
the sample is within the normal range experienced by samples in all major sonication instruments
(please see referenced publication on page 9).
Only use room temperature (not previously frozen) 96-Well PIXUL Plates (cat. no. 53139) with the
PIXUL instrument. Place plates with well A1 in the upper left corner. All wells in columns being
sonicated MUST be filled with fluid. Do not reuse plates from partial runs, as this can lead to issues
with shearing consistency and sample contamination.
Don't lift the external lid during processing of samples. Lifting the external lid initiates drainage
of PIXUL Coupling Fluid from under the plate, necessitating re-circulation of Coupling Fluid to
continue processing samples.
Any technical modifications to the device are prohibited. Service works may be carried out only by
qualified personnel.
Personal Protective Equipment, such as long sleeve clothing, safety footwear, protective gloves,
and protective goggles, is required for certain activities.
www.activemotif.com
1