DAW Industries 4P10 KD Manual - Page 4

Browse online or download pdf Manual for Medical Equipment DAW Industries 4P10 KD. DAW Industries 4P10 KD 10 pages.

2
True-KD
Series
Knee Connection to KD Adapter (TSC-KDL)
The instructions below apply to True-KD
using the KD Adapter (TSC-KDL):
After lamination of the KD Adapter (TSC-KDL) is complete,
A. Use the included
7
/
Bolt to remove the black Spacer from the
16
center of the top of the knee .
Spacer
7
/
Bolt
16
B. Remove the Kneecap .
C. Flex the 4P10 KD
True-KD Knee approx . 90-degrees .
D. Select your desired degree of external knee rotation .
Each notch of the KD Adapter and knee top is an adjustment
of 2 degrees .
E. Secure the knee to the KD Adapter using the included
7 /
16 Bolt (8mm hex) . Torque the bolt to 9ft-lb (12.2Nm).
DAW
®
DAW
Industries
© Copyright 2021, DAW Industries, San Diego, CA - All Rights Reserved 01082021
Series Knees, when
Spacer
7
/
16
Lock Washer
7
/
Bolt
16
(8mm hex)
For Technical Support call (800)242-8669
Bolt
DAW
®
DAW
Industries
For Technical Support call (800)242-8669
© Copyright 2021, DAW Industries, San Diego, CA - All Rights Reserved 01082021
15
Patient Notes