6.

Installation Guide:

1. Place the Universal Therapy System (UTS) directly on to bed frame and ensure CPR
valve is at the head end of the bed. Secure and check that the bed straps on the underside
of mattress are fastened to the bed frame correctly (Check moving sections on the bed
frame to make sure that they are not encumbered by the straps and can continue to
operate normally).
2. Hang the handset on the bed rail or mattress platform and plug the power cord into mains
outlet. Ensure that electrical cables are safe, tidy, free of obstruction and are not able to
be caught in the bed frame. The power LED will be at standby mode with amber light
illuminating.
CAUTION: it is important to routinely inspect power cable to ensure it is not
obstructing or causing a tripping hazard. Check and make sure that the power
cable is not under strain or damaged.
3. Ensure that the CPR valve is set to "CLOSE" position (Fig. 1).
4. Turn the power on from handset (Green LED will illuminate).
5. When the Universal Therapy System (UTS) is turned on, the pump will start to inflate the
mattress for around 30 minutes in auto firm mode. During this inflation period, the "AF"
LED will be displayed on LED window to indicate the mattress is not ready to use. When
"AF" switches to "AL" with display on LED window, bed sheet can be added (NOT tucked
under mattress, left to hang freely) and the patient can be placed and positioned on the
mattress surface. The audible alarm is disabled during the inflation period and will resume
its function when "AF" LED goes off.
6. Once the mattress is properly inflated, system will switch to alternating mode. Press "Auto"
button to enable auto detection feature and the system will automatically set to an optimum
pressure according to patient's weight.
The Universal Therapy System (UTS) also allows the caregiver the flexibility to disable
"Auto" feature and switch to manual adjustment, by repeatedly pressing "LEVEL" button to
accommodate an individual's comfort need. Default manual setting will be the level
detected during "Auto" mode, or at Level 2 (02 displayed on LED window) if Auto detection
is incomplete.
Please Note: "Auto Firm (AF)" and "STATIC (St)" mode is designed to return to
"Alternating (AL)" mode after 30 minutes of activation.
Fig. 1 Turn the dial to CLOSE
position to inflate air mattress.