Cleaning
During use, the meter should be kept free
of liquid to ensure drying does not occur
inside the meter. If drying should occur,
the nutator will stick or drag, affecting ac-
curacy. If this occurs, cleaning is required.
Cleaning the Nutator
1. Loosen the screws on the back of
the meter about 1/4 inch (0.6 cm).
(Figure 4)
2. Tap the screws with a hammer to
loosen the coverplate from the hous-
ing. (Figure 5) Remove the screws
and washers.
3. Gently pry the coverplate from the
housing. Take care not to damage
the coverplate, housing, or O-ring.
(Figure 6)
10
4. Remove the signal generator from
Figure 4
the pin on the nutator disc.
(Figure 7)
5. Gently remove the metal clips holding
the nutator in place. (Figure 8) Take
Figure 5
care not to damage the O-ring on the
nutator inlet.
6. Remove the nutator.
7. Using water or a solvent-based fluid
and a brush, carefully remove residue
from the nutator disc, nutator halves,
signal generator, and inside of hous-
ing. Allow all parts to dry.
Figure 6
Figure 7
Figure 8