Gait up Physilog 5 User Manual - Page 2

Browse online or download pdf User Manual for Analytical Instruments Gait up Physilog 5. Gait up Physilog 5 12 pages.
Also for Gait up Physilog 5: User Manual (17 pages), User Manual (17 pages)

Gait up Physilog 5 User Manual
1. Physilog®5 elements

2. Charging

The Physilog®5 is charged via the microUSB port. Plug the microUSB connector of the cable
delivered with the Physilog®5 to the Physilog®. Plug the USB connector to a Computer or
Charger (see voltage specifications below).
Warning: Be careful to plug the microUSB cable in the correct orientation into the Physilog®,
forcing the connector may permanently damage the Physilog® and compromise its
waterproofing.
Low battery level is indicated by orange blinking during measurement or if the Physilog®
blinks rapidly 3 times red when starting and doesn't start the recording (see LED indications
below). A battery level estimation is available in the mobile application. When the Physilog®
is discharged, plug it for charging for at least 30 minutes before use.
It is recommended to avoid complete discharging of the Physilog®5, therefore check that the
Physilog® sensors are off when you store them. After a long period without using the
Physilog®, plug it for charging for at least 30 minutes before use. It may happen that the LED
doesn't blink orange at the beginning of charging after the battery was very low, to see the
charging ongoing shortly push the main button or disconnect and reconnect after some
charging time.
Supply Voltage specifications
The Physilog® 5 should be charged using USB at 4.2V, 100mA. For charging, the ambient
temperature must lie between 0°C and 45°C.
4 jan.2017 - Physilog 5 user manual.pdf - v1.1 - Check www.gaitup.com for the latest version. Copyright 2017 Gait Up SA. CH-1015 SWITZERLAND