6.3. Technical Error

Errors are indicated by a flashing LED light on the start button. They could be reset by pressing the start button or
restarting the device. If the error continues, the LED light will flash again. The device does only work after eliminating
the error.
Error
The device is not running,
no LED light
Flashing LED
Rapidly Flashing LED
1 x Flashing LED
2 x Flashing LED
Flash Mode: ON 2000ms / OFF 300ms / ON 300 ms / OFF 300ms / ON 2000ms
3 x Flashing LED
Troubleshooting
Defective fuse
ON/OFF switch in off-state
The lid has not been opened
after switching on the device (with a
closed lid)
The lid has not been opened before
the further mixing process
The device cover does not close
properly after pressing the start
button
Flash Mode: ON 300ms / OFF 300ms, 1.6 Hz
Device activated for more than
recommended on-time
Overheating
Flash Mode: ON 100ms / OFF 100ms, 5 Hz
Defective locking
Missing locking signal
Flash Mode: ON 2000ms / OFF 300ms / ON 2000 ms
Defective incremental decoder
The defective motor control unit
Flash Mode: ON 2000ms / OFF 300ms / ON 300 ms / OFF 300ms /
ON 300 ms / OFF 300ms / ON 2000ms
Check the fuse and replace it
if necessary
Activate the ON/OFF switch
The lid circuit is not yet activated. Open
and close the lid
Open and close the lid
Close the lid, press the start button
Let the device cool down
Let the device cool down
Let a service technician
check the lock
Let a service technician
check the lock
Request service technician
Replace motor and control (service
technician)
11
Solution