Annex II. IQ/OQ

Functional Test of the gako PM140
Conclusion
Assessment
OK
Not OK
Check the device label for readability.
Check device location for stability and suitability.
Check the device location for drafts, direct sunlight, extreme humidity, or temperature
fluctuations.
Check the device location for cleanliness, dryness, and freedom from dust.
Check the accessibility to the power plug and power switch.
Check the power supply for suitability and country-specific nominal voltage and frequency.
Handling by unauthorized persons is excluded.
Operation instructions in the user manual are available.
Turn the device upside down and remove the safety screws from the bottom.
Turn the device back to its rubber feet and open the lid.
Remove the cardboard tube from the device and close the lid.
Connect the device to the local power supply using the provided connection cable.
Switch the power button on the back side of the device to "I".
Check if the red light keeps flashing after turning it on.
Prepare a PM jar adding a transparent gel and close the lid of the jar to test deaeration (both
gako PM jar 100ml HV and gako PM jar 125ml HV+LV can be used).
Place the PM jar holder on the table, insert the PM jar into the PM jar holder, and turn the jar
slightly clockwise. It is normal to be tight.
Is the PM jar inserted in the PM jar holder?
Place the PM jar with the PM jar holder into the well. Make sure the PM jar lid is completely
closed, the shape is complementary with the well and the jar is pushed down.
Is the PM jar fully inserted into the well?
Close the lid of the device and set the time to 30 seconds from the button.
Does the button move properly?
Press "START".
Does the device start operation?
Is the operation time the same as the set time?
Does the device stop operating automatically?
Open the lid of gako PM140 and remove the PM jar with the PM jar holder.
Put the PM jar holder on the table.
Turn the PM jar slightly clockwise and take it out (it is normal to be tight).