Autonics KRN1000 Series Instruction Manual
Browse online or download pdf Instruction Manual for Measuring Instruments Autonics KRN1000 Series. Autonics KRN1000 Series 2 pages. Lcd touchscreen paperless recorder
Also for Autonics KRN1000 Series: User Manual (15 pages), Instruction Manual (2 pages)
AEP-E-0180C
LCD Touchscreen Paperless Recorder
KRN1000 SERIES
I N S T R U C T I O N
M A N U A L
Thank you for choosing our Autonics product.
Please read the following safety considerations before use.
Safety Considerations
※Please observe all safety considerations for safe and proper product operation to avoid hazards.
※Safety considerations are categorized as follows.
Warning
Failure to follow these instructions may result in serious injury or death.
Caution
Failure to follow these instructions may result in personal injury or product damage.
※The symbols used on the product and instruction manual represent the following
symbol represents caution due to special circumstances in which hazards may occur.
Warning
1. Fail-safe device must be installed when using the unit with machinery that may cause serious injury or substantial economic
loss. (e.g. nuclear power control, medical equipment, ships, vehicles, railways, aircraft, combustion apparatus, safety
equipment, crime/disaster prevention devices, etc.)
Failure to follow this instruction may result in personal injury, fire, or economic loss.
2. The unit must be installed on a device panel before use.
Failure to follow this instruction may result in electric shock.
3. Do not connect, repair, or inspect the unit while connected to a power source.
Failure to follow this instruction may result in electric shock .
4. Do not disassemble or modify the unit. Please contact us if necessary.
Failure to follow this instruction may result in electric shock or fire.
5. Check the terminal numbers before connecting the power source.
Failure to follow this instruction may result in fire or burning the unit.
Caution
1. Do not use the unit outdoors.
Failure to follow this instruction may result in shortening the life cycle of the unit, or electric shock.
2. Use the unit within the rated specifications.
Failure to follow this instruction may result in shortening the life cycle of the unit, or fire.
3. Do not use water or oil-based detergent when cleaning the unit. Use dry cloth to clean the unit.
Failure to follow this instruction may result in electric shock or fire.
4. Do not use the unit where flammable or explosive gas, humidity, direct sunlight, radiant heat, vibration, or impact may be
present.
Failure to follow this instruction may result in fire or explosion.
5. Keep dust and wire residue from flowing into the unit.
Failure to follow this instruction may result in fire or product damage.
6. Check the polarity of the power contact before wiring the unit.
Failure to follow this instruction may result in fire or explosion.
※Please read "Safety Considerations" in KRN1000 user manual before using.
Ordering Information
KRN1000
04
0
1
0
S
Case
S
Standard panel installation type
Power supply
0
100-240VAC 50/60Hz
Communication output
RS422/485+Ethernet+USB Device
1
0
None
1
Alarm relay output 8 channels
2
Alarm relay output 6 channels + Digital input 2 channels
Option input/output
Alarm relay output 6 channels
3
+ 24VDC power output for transmitter
Alarm relay output 4 channels + Digital input 2 channels
4
+ 24VDC power output for transmitter
04
4 channels
Input channels
08
8 channels
12
12 channels
16
16 channels
Item
KRN1000 Paperless recorder
User Manual
For the detail information and instructions, please refer to user manual and user manual for communication.
Download manuals from the enclosed CD-ROM or our web site (www.autonics.com).
※The above specifications are subject to change and some models may be discontinued without notice.
Specifications
Series
KRN1000
Power supply
100-240VAC 50/60Hz
Allowable voltage range
85 to 110% of rated voltage
Power consumption
Max. 23VA
Display method
5.6 inch TFT Color LCD
Resolution
640×480 pixels
Screen
Adjusting brightness
3-level (Min/Standard/Max)
Input method
Touch screen (Pressure sensitive type)
Number of input channels
4 / 8 / 12 / 16 channels
※1
Universal input
Temperature sensors (thermocouple, RTD), Analog (voltage, current (shunt))
1 to 4-CH: 25ms/125ms/250ms, 5 to 16-CH: 125ms/250ms
Sampling period
(internal sampling period is average movement filter and alarm output operation unit time)
Recording period
1 to 3600 sec
Internal memory
Approx. 200MB
External memory
SD / USB memory max. 32GB
2300VAC 50/60Hz for 1 min (between power terminals and case)
Dielectric strength
※Except ethernet and USB device
Mechanical
10 to 60Hz 4.9m/s
in each X, Y, Z direction for 1 hour
2
Vibration
Malfunction
10 to 60Hz 1m/s
in each X, Y, Z direction for 10 min
2
Insulation resistance
Over 20MΩ (at 500VDC megger)
Noise immunity
Square shaped noise by noise simulator (pulse width 1µs) ±2kV
Time accuracy
Within ±2 min/year (available up to 2099)
Protection structure
IP50 (front part)
Ambient temperature
0 to 50℃, storage: -20 to 60℃
Environ-
ment
Ambient humidity
35 to 85%RH, storage: 35 to 85%RH
Approval
Weight
※2
Approx. 1290 to 1400g (approx. 590 to 700g)
※1: For more information of universal input, refer to '
Input/Output'.
※2: The weight includes packaing. The weight in parenthesis is for unit only.
※Environment resistance is rated at no freezing or condensation.
Unit Description
Front part
1. Power indicator: Power turns ON and the red LED turns ON.
1
2. Screen: Measrued value is displayed as trand graph, bar bar graph,
digital figures.
3. Front cover: Open the front cover. There are power switch and, USB
2
Host/Device, SD card slot.
4. Power switch: Turn ON/OFF the power of KRN1000.
5. USB host port: Connect the USB memory.
It recognizes up to 32GB. When using extension cable,
cable length should be up to 1.5m.
6. USB device port: Used for parameter settings.
7. SD card slot: SD card memory slot. It supports up to 32GB.
8. Stylus pen: Used for touching screen.
※Do not connect the other USB devices except USB memory at USB host
4
port.
7
5
6
8
3
Rear part
1. Sensor input terminal: Connects universal input.
2. Ethernet port: Connector for ethernet cable.
It communicates Modbus TCP.
5
3. RS422/485 port: Connects RS422/485 for Modbus RTU communication.
1
4. Option input/output port: Connects for option input/output (digital
5. Power input: Power connection (100-240VAC 50/60Hz)
4
3
2
Dimensions
144
11.5
● Bracket
69.2
15
Input/Output
Type
Input/Output type
Input type
Universal input
Input impedance
Display
accuracy
Resolution
Digital
input
Option
Alarm relay
input/output
※3
output
Power output for transmitter
RS422/485
Communication
Ethernet
output
※5
USB Device
※1: Current measurement and connection examples
Connect 250Ω shunt resistance and set analog input type 0-20mA (shunt) / 4-20mA (shunt).
It is available to measure 0-20mA / 4-20mA current.
※2:
At room temperature (23℃ ± 5℃)
RTD Cu50Ω (-200≤T≤200): (±0.1% F.S. or ±1.5℃, select the higher one) ±1-digit
RTD DPt50Ω (-200≤T≤500): (±0.1% F.S. or ±1.5℃, select the higher one) ±1-digit
Thermocouple R, S, C, G type (0≤T≤100): (±0.1% F.S. or ±4.0℃, select the higher one) ±1-digit
Thermocouple U, T type (-100≤T≤400): (±0.1% F.S. or ±2.0℃, select the higher one) ±1 digit
Thermocouple B type, below 400℃: There is no accuracy standards.
All thermocouples, below -100℃: (±0.3% F.S. or ±4.0℃, select the higher one) ±1-digit
Out of room temperature range
RTD Cu50Ω (-200≤T≤200): (±0.2% F.S. or ±3.0℃, select the higher one) ±1-digit
RTD DPt50Ω (-200≤T≤500): (±0.2% F.S. or ±3.0℃, select the higher one) ±1-digit
※3: Input/Output is different by option. Refer to '
※4: For supplying power for transmitter, it is recommened to use shield cable to reduce noise.
※5: RS422/485, ethernet, USB device communication outputs are not used at the same time.
※If sensor input line is longer, it is recommended to use shield cable to reduce noise.
Input/Output Circuit
Universal input
Option input/output 1
(alarm output 8 channels)
Option input/output 3
(alarm output 6 channels + power for transmitter output)
input (non-contact/ contact), alarm output,
power for transmitter).
Comprehensive Device Management Program [DAQMaster]
DAQMaster is the comprehensive device management software for setting parameters and monitoring processes.
DAQMaster can be downloaded from our web site at www.autonics.com.
Item
Minimum specifications
System
IBM PC compatible computer with Pentium Ⅲ or above Hard disk
Operations Windows 98/NT/XP/Vista/7/8/10
Memory
256MB+
(unit: mm)
Error Message
● Panel cut-out
Min. 175
Error message
Descriptions
M4
When input type is temperature sensor (Thermocouple, RTD) and the measurement value is higher than high-limit value of input
range, it flashes HHHH. It is cleared when the measurement value is within the high-limit range.
HHHH
When input type is analog (voltage, current (shunt)) and the measurement value is over 10% of high-limit input range, it flashes
HHHH. It is cleared when the measurement value is within 10% of high-limit input range.
14
When input type is temperature sensor (Thermocouple, RTD) and the measurement value is lower than low-limit value of input
range, it flashes LLLL. It is cleared when the measurement value is within the low-limit range.
LLLL
When input type is analog (voltage, current (shunt)) and the measurement value is over 10% of low-limit input range, it flashes
138
+1.0
0
LLLL. It is cleared when the measurement value is within 10% of low-limit input range.
When input type is temperature sensor (Thermocouple, RTD) and input is break, it flashes BURN.
BURN
It is cleared when input is connected.
When forgetting and entering unvaild password 3 times, "ASKey" appears with error message.
ASKey
Contact our service center with ASKey.
Description
RTD
JPt100Ω, DPt100Ω, DPt50Ω, Cu100Ω, Cu50Ω (supplied current: approx. 190µA)
Thermocouple
B, C (W5), E, G, J, K, L, L (Russia), N, P, R, S, T, U
Voltage: ±60mV, ±200mV, ±2V, 1-5V, ±5V, -1V-10V
Analog
Current: 0-20mA, 4-20mA (measureable when using 250Ω shunt resistance)
Voltage (V): Approx. 205kΩ
RTD, Thermocouple, Voltage (mV): Min. 200kΩ
RTD
Warm-up time: Max. 30 min
Thermocouple
At room temperature (25±5℃): ±0.1% F.S.±1-digit
※2
Out of room temperature: ±0.2% F.S.±1-digit
Analog
16-bit
No-contact input
ON: Residual voltage max. 1V, OFF: Leakage current max. 0.1mA
Contact input
ON: Max. 1kΩ, OFF: Min. 100kΩ, Short-circuit: Approx. 4mA
Capacity
250VAC 3A, 30VDC 3A, 1 Form A (resistive load)
Mechanical: Min. 20,000,000 operations
Life cycle
Electrical: 100,000 operations (3A 250V AC, 3A 30V DC)
24±2VDC, Max. 60mA ※Built-in over current protection circuit
※4
Modbus RTU ※It is recommended to use shielded cable over AWG 24.
IEEE802.3 10 BASE-T / IEEE802.3U 100 BASE-TX (Modbus TCP)
USB V2.0 Full Speed (Modbus RTU)
Ordering Information'.
Communication output
Option input/output 2
(alarm output 6 channels + digital input 2 channels)
Option input/output 4
(alarm output 4 channels + digital input 2 channels
+ power for transmitter output)
Item
Minimum specifications
1GB+ of available hard disk space
VGA
Resolution: 1024×768 or higher
Others
RS232C serial port (9-pin), USB port
※1