ROUTINE MAINTENANCE

The LifeChoice Activox POC does not have any end-user repairable parts.
Fan Inlet Vent Cleaning: There is one inlet vent located on the right
side (facing the unit) of the LifeChoice Activox POC. The carrying case
for the unit has mesh material corresponding to the location of the fan
inlet vent. You should visually check the carrying case to ensure that
there is no build-up of lint, hair or other materials that could obstruct
the flow of air into the unit's fan inlet vent. Remove any material build-up
using a soft brush.
Unit and Carrying Case Cleaning: If it is necessary to clean the unit
and/or carrying case, use only warm water and a mild liquid dish
detergent. Dampen a cloth in the solution and carefully wipe the
outside surfaces of the unit and/or carrying case.
WARNING: DO NOT use alcohol, solvents, polishes or any oily
!
substances since they may be flammable.
Nasal Cannula Inspection: At least once a week, visually inspect the
nasal cannula. Make sure there are no kinks or obstructions. Replace
as needed with a standard 7-foot (2-meter) single-lumen type cannula
from your local medical equipment supplier.
Reserve Oxygen Supply: Your equipment provider should provide
or suggest an alternative source of supplemental oxygen therapy in
case there is a mechanical failure or power outage.
Hour Meter: The hour meter function is for use by trained service
personnel.
16
At least once a month inspect the
grille of the intake vent and if it appears
dirty, use a dry cotton swab (Q-Tip®
or similar) to clean it. Never use
alcohol, sol vents, or p et roleu m
distillates for cleaning.

SPECIFICATIONS

Maximum outlet pressure: 19 psi (131 kPa)
Device Electrical Classification: Class II, Type BF
Oxygen Concentration: 90% (+/- 3%)*
Oxygen Concentration Sensor Alarm: 85% or less
Flow Control Pulse Settings: 1, 2, & 3 LPM (Equivalent to
Continuous Flow) 20%
Mode of Operation: Continuous equipment and mobility: portable
Weight: 4.83 lbs.
Dimensions: 9.05" Width x 7.875" Height x 4.38" Depth
Power Requirement: AC 100 to 240 VAC, 50 to 60 Hz;
DC 12 to 19 VDC or Battery 12-16.8 VDC
LifeChoice Activox POC Input: 19VDC, 6A
External AC Power Adapter/Charger, XYC103AC-Input:
100-240VAC, 2.5A; Output: 19V, 6.3A
External DC Power Adapter/Charger, XYC104DC-Input:
11.5-16VDC, 10A; Output: 19V, 6.3A
Approximate Internal Battery Duration: See Battery Life Table on
following page
Minimum Internal Battery Recharge Time: 4 Hours
Sound: At 10 Feet (3m) - 41 dB at 1 LPMeq, 44 dB at 2 LPMeq,
46 dB at 3 LPMeq
Environmental Conditions for Use, Storage and Transport:
Operational Temperature: 41°F to 104°F (5°C to 40°C)
Operational Humidity Range: 15% to 95%, Non-condensing
Operational Altitude Max: 10,000 Feet (3km) or 10.1 PSIA (522 mmHg)
Storage Temperature: 32°F to 140°F (0°C to 60°C)
Storage Humidity Range: Up to 95%, Non-condensing
* When powered by the external AC adapter, external DC adapter, or external or internal battery.