dynarex 10405 User Manual - Page 7
Browse online or download pdf User Manual for Medical Equipment dynarex 10405. dynarex 10405 10 pages. Bariatric hd full electric home care bed
Also for dynarex 10405: Manual (16 pages)
Cleaning — It is recommended that the following
cleaning procedures be conducted between uses
and at least once a year:
1. Unplug the power code of electric beds before
performing maintenance.
2. Wipe down with disinfectant. Allow to dry completely.
CAUTION: Avoid spraying water directly into the elec-
tric components to avoid damage.
Mechanical Inspection and Maintenance:
1. Inspect all bed components for damage or excessive
wear.
2. Inspect all mounting hardware for bending, damage
or excessive wear.
3. Inspect all bolts and rivets to ensure that they are
securely tightened and functioning properly.
4. Inspect caster wheels to ensure they roll properly.
Electrical Inspection and Maintenance:
1. Inspect all electrical components for damage or
excessive wear.
2. Check pendent, power code and motor cables for
chafing, cuts or excessive wear.
3. Make sure all of plugs are fully attached and free of
damage.
4. Check all functions working properly.
Bariatric HD Full Electric Home Care Bed (Reorder Nos. 10403 & 10405)
PREVENTATIVE MAINTENANCE
To Maximize Services Life— To ensure maximum life
of your bed, follow all of warning and cautions in
the User Manual and maintain your bed with care—a
thorough inspection should be conducted as follow:
1. Run the bed frame to either the highest or lowest
position once a day.
2. Observe bed operation and report any discrepancies
or problems to service personnel.
3. Check and tighten all tubes and mounting hardware
monthly.
4. Check and tighten all electric wires and cables are
properly connected monthly.
5. Lubricate pivot point, pins and bolts every 2 months.
WARNING: Possible Shock Hazard. All electric compo-
nents to ONLY be replaced by service personnel.
7