Display reads an
abnormal result
Bluetooth
connection
unstable
No response
The cuff does not
inflate properly or
pressure falls
quickly during test
Irregular heartbeat
(arrhythmia)
The cuff was not
properly applied
The cuff position
was not correct or
it was not properly
tightened.
Body posture was not
correct during
testing
Speaking, moving
arm or body, being
angry, excited or
nervous during test
Bluetooth
connection
unsuccessful,
monitor is
abnormal, or strong
electromagnetic
interference is
present
Incorrect operation
or strong
electromagnetic
interference
Review the cuff application
instructions and retest
It is inappropriate for
people with serious
arrhythmia to use this
monitor.
Check with physician.
Review the cuff application
instructions and retest
Review the cuff application
instructions and retest.
Review body posture
instructions and retest
Retest when calm; avoid
speaking or movement during
the test
Reset iOS device. Reset
monitor by pressing the
START/STOP button about 10s.
Make sure the monitor and iOS
device are away from other
electrical equipment. Please
see GENERAL SAFETY AND
PRECAUTIONS.
Press the START/STOP button
about 10 seconds to reset the
device, relaunch app, and
reconnect the iOS device to
the monitor