Emerson Rosemount 5408 Quick Start Manual - Page 6

Browse online or download pdf Quick Start Manual for Transmitter Emerson Rosemount 5408. Emerson Rosemount 5408 20 pages. Process seal antenna

Emerson Rosemount 5408 Quick Start Manual
Quick Start Guide
Step 2: Tighten the bolts and nuts
Note
Re-tighten after 24 hours and again after the first temperature cycle.
Check at regular intervals and re-tighten if necessary.
Table 1. Torque Value, lb-ft (N-m)
Process
connection size
2-in./DN50/50A
3-in./DN80/80A
4-in./DN100/100A
1. The conditions used for the calculation are: Standard mating metal flange, A193 B8M Cl.2 / A4-70
bolt material, and a friction coefficient of μ=0.16.
2. Low strength bolt and non-metallic mating flange may require lower tightening torque.
Step 3: Align the transmitter head (see page 8)
6
(1)(2)
ASME B16.5
Class 150
Class 300
29 (40)
52 (70)
33 (45)
48 (65)
59 (80)
52 (70)
Process connection rating
EN1092-1
PN6
PN10/PN16
15 (20)
26 (35)
37 (50)
37 (50)
37 (50)
37 (50)
October 2017
JIS B2220
PN25/PN40
10K
29 (40)
18 (25)
41 (55)
22 (30)
74 (100)
26 (35)