DCI Edge SERIES 5 Installation Instructions Manual - Page 6

Browse online or download pdf Installation Instructions Manual for Medical Equipment DCI Edge SERIES 5. DCI Edge SERIES 5 13 pages. Side delivery system
Also for DCI Edge SERIES 5: Installation Instructions Manual (12 pages), Installation Instructions Manual (11 pages), Instructions For Use Manual (8 pages), Installation Instructions Manual (14 pages), Instructions For Use Manual (19 pages)

DCI Edge SERIES 5 Installation Instructions Manual
SERIES 5 DENTAL CHAIR
Installation Instructions

CHAIR PREPARATION

SHIPPING BOLT
Remove the shipping bolt by unscrewing from
above the seat pan casting using a 3/8" wrench.
Once unscrewed, removed the bolt and related
hardware from the assembly.
POWER SWITCH COVER
Power Switch Cover: Install the power switch
cover into the opening around the power switch
by snapping it firmly into place.
FOOT SWITCH
Foot Switch: Unpack the foot switch and set in
position for use.
BACKREST
Rotate the backrest into position by aligning
1
the backrest bottom holes to the holes in the
link bar which extends from the upper hydraulic
assembly.
6
RETURN TO LAST POSITION
PRESET POSITION "0"
Loosen the set screw which secures the steel
2
cross pin. Align the holes of the backrest and
the back link and re-insert the cross pin. If neces-
sary, use a rubber mallet to drive the pin until it has
fully engaged both parts.
Tighten the set screw on the backrest link
3
against the pin to prohibit its movement.
PRESET POSITION "1"
FOOTPAD