Maico MA 28 Quick Manual - Page 2
Browse online or download pdf Quick Manual for Measuring Instruments Maico MA 28. Maico MA 28 2 pages.

Explanation of the Function Keys (F1-F4)
Menu
Function
screen
Button
F1
F2
st
1
menu
screen
F3
F4
F1
nd
2
menu
F2
screen
F3
F4
F1
rd
3
menu
F2
screen
F3
F4
F1
F2
th
4
menu
screen
F3
F4
4 TEST PREPARATION
Step 1: Preparing the patient for a test
✓ Put the patient at ease and instruct on the test procedure:
"I'm going to place these headphones on your ears. You will hear some beeping sounds. Press the patient
response switch (or raise your hand) as soon as you hear the beep, even if it is very soft."
✓ Have the patient sit at an angle with his/her back to the tester/audiometer, so that the patient cannot see
the adjustments made on the audiometer.
Step 2: Place the headphones on the patient
✓ Place the headphones on the patient with the RED phone on the RIGHT ear and the BLUE phone on the
LEFT ear. Make sure the headband sits snuggly on the crown of the head and the earphones are centered
over both ears.
Step 3: Familiarize the patient with the test procedure
✓ Familiarization with the test procedure is recommended prior to starting the actual test. This is typically done
by starting in the better hearing ear at 1000 Hz or higher and at an elevated level (i.e. dB HL).on with the
test procedure is recommended prior to starting the actual test. level (i.e. dB HL).
5 MANAGING TEST RESULTS
There are three options to manage test results stored on the device:
• Save results to the internal database.
• Print results (optional printer required)
• Transfer result to the PC software (optional MAICO Sessions software required).
For further information ask your local distributor or MAICO Diagnostics GmbH, Sickingenstr. 70-71, 10553 Berlin, Germany
Tel.:+49 30 70 71 46-50 • Fax: +49 30 70 71 46-99 • E-mail: [email protected] • Internet: www.maico.biz
8507409 Rev. 2
Label
Description
Air
Selection between Air or Bone conduction transducer.
Bone
Right
Selection between Right or Left ear
Left
Steady
Pulse
Type of signal presented. Options include Steady, Pulse, Warble, Pulse
and Warble (P&W).
Warble
P&W
More
To move to the 2
To delete single stored measurement. Set device with the ear, test type and
Del
frequency to delete the measurement.
New
To delete all stored results for starting a new test.
Print
To print test results. (Optional printer required)
More
To move to the 3rd menu screen.
HW
Activates Hughson-Westlake automatic air conduction threshold test.
Save DB
To save results to internal database.
View DB
To view results saved to the internal database.
More
To move to the 4
-
No function
-
No function
Menu
To move to Settings menu.
Back
Returns the function key selection to the 1
nd
menu screen.
th
menu screen.
2 of 2
MA 28 – Quick Guide
st
menu screen.
11/05/2020