Dynisco 1490 Operating Manual - Page 18
Browse online or download pdf Operating Manual for Touch Panel Dynisco 1490. Dynisco 1490 20 pages. Din indicator
Also for Dynisco 1490: Product Manual (2 pages)
![Dynisco 1490 Operating Manual](/uploads/products/server2/1994241/1994241.jpg)
From lab to production,
providing a window into the process
OUTPUTS
Single Relay
Contact Type &
Rating:
Lifetime:
Isolation:
Dual Relay
Contact Type &
Rating:
Lifetime:
Isolation:
Linear DC
Accuracy:
Resolution:
Isolation:
Transmitter PSU
Power Rating:
Isolation:
SERIAL COMMUNICATIONS (RS485) (option)
Physical:
Protocols:
Isolation:
OPERATING CONDITIONS (FOR INDOOR USE)
Ambient Temperature:
Relative Humidity:
Altitude
Supply Voltage and
Power:
Single pole double throw (SPDT), latching or non-latching action
(selectable); 2A resistive at 120/240VAC.
>500,000 operations at rated voltage/current.
Basic Isolation from universal input and SSR outputs.
Single pole single throw (SPST), latching or non-latching action
(selectable); 2A resistive at 120/240VAC.
>200,000 operations at rated voltage/current.
Reinforced safety isolation from inputs and other outputs.
±0.1% of span (mA @ 250Ω, V @ 2kΩ).
15 3/4 bit (1 part in 52K) and updated at approx. 65ms intervals.
(130ms settling time)
Reinforced safety isolation from inputs and other outputs.
24V Tx PSU Module; Unregulated 18 to 32V DC into 400 Ω min
Linear Output Module; Regulated 0.0 to 10.0V into 500 Ω min.
Reinforced safety isolation from inputs and other outputs.
1200, 2400, 4800, 9600 or 19200 bps.
Selectable between Modbus and West ASCII.
Reinforced safety isolation from all inputs and outputs.
0°C to 55°C (Operating), –20°C to 80°C (Storage).
20% to 95% non-condensing.
<2000m
100 to 240VAC 10%, 50/60Hz, 9VA
(for mains powered versions), or
20 to 48VAC 50/60Hz 9VA or 22 to 65VDC 5W
(for low voltage versions).
P/N: n/a
Rev: n/a
www.dynisco.com
-18-
ECO: n/a