Relative humidity (non-
condensing)
Barometric
Ingress rating
Drop test
Physical
Size
Device
Charge/Case
Packing size
Total Weight
(Device+Case+Package)
Connector
Power Supply
Battery type
Battery run time (full charge)
Charge time
Cycle life
ECG
Lead type
Lead set
Sampling rate
Sampling accuracy
Display Gain
Bandwidth
Electrode offset potential
tolerance
HR measurement range
HR accuracy
Measurement summary
For support, please contact at +1-650-397-6103 or visit www.dynosense.com
All manuals and user guides at all-guides.com
10% to 95%
10% to 95%
800 to 1060
800 to 1060 hPa
hPa
IPX4
1.0 m
65.13mm X 42.35mm X 97.08mm
106.69mm X 126.98mm X 68.39mm
152mm X 89mm X 132mm
~600 g
Micro USB Connector
Rechargeable Lithium-Polymer
battery 210 mAh
Device usage twice daily: ~1 month
Standby Mode: ~2 weeks
In-case Mode: ~3 months
<3 hrs. to 90%
>300 times
Single Lead, 3 Contacts
Lead I
500 Hz
24 bit
10 mm/mV
0.67 to 40 Hz
±300mV
30 to 250 bpm
±2 bpm or ± 2% (of reading),
whichever is greater
Heart Rate
SpO
2
Standard Confromance
SpO
display range
2
SpO
Accuracy (A
2
Sampling Rate
PR display range
PR accuracy
Measurement summary
Wavelength / Max emission
power
Ther mometer
Technique
Environment temperature
Measurement site
Measurement range
Accuracy
Respiration Rate
Technique
Measurement range
Accuracy
16
For support, please contact at +1-650-397-6103 or visit www.dynosense.com
Meet standards of ISO 80601-2-61
0% to 100%
)
Range
Accuracy
r ms
70% to 100 %
±2 %
250Hz
30 to 250 bpm
±2 bpm or ±2% of the reading,
whichever is greater
SpO
, PR
2
660 nm/905 nm, 1.96 mW/1.10 mW
Ther mo-resistive
10.0°C to 40.0 °C
Oral, under tongue
30.0 °C to 43.0 °C
± 0.2 °C
HR V based
8 bpm – 30 bpm
± 2bpm or ± 2% of the reading,
whichever is greater
17