Decagon Devices ECH2O EA-10 Integrator Manual - Page 6
Browse online or download pdf Integrator Manual for Accessories Decagon Devices ECH2O EA-10. Decagon Devices ECH2O EA-10 14 pages. Soil moisture sensor
EA-10 Soil Moisture Sensor Integrator's Guide
2. Installing the EA-10
The EA-10 monitors the water content of the soil in
which it is placed by measuring the dielectric constant
of the soil and water surrounding the probe. While the
probe measures just the soil which is adjacent to it, its
reading is most useful if that measurement represents
the general soil conditions in which the probe is
placed. For that to happen, the monitoring site must
be carefully sel-ected, and the probe must be properly
installed.
Procedure
When installing the ECH
mize contact between the probe and the soil. There
are two methods to accomplish this.
1. Use Decagon's Probe Installation Kit to install the
probe. This kit has a custom-shaped blade to make
the insertion in the soil, then another tool to place
the probe into the insertion. For deeper installa-
tions, use an augur to reach the desired depth,
then use the Installation kit with extension rods to
install the probe.
2. Use a thin implement like a trenching shovel, gar-
dening spade, or flat bar to make a pilot hole in the
soil. Then insert the probe into the hole, making
sure the entire length of the probe is covered.
Finally, insert the shovel again into the soil a few
inches away from the probe, and gently force soil
toward the probe to provide good contact
2. Installing the EA-10
O probe, it is best to maxi-
2
4