Gamma 700K Instruction Manual - Page 7

Browse online or download pdf Instruction Manual for Blood Pressure Monitor Gamma 700K. Gamma 700K 14 pages. Aneriod blood pressure kit

indicator zone. If the needle points outside of the accuracy
indicator zone, the gauge will give inaccurate readings. In
this case you have to bring the device to the nearest service centre
for calibration (address mentioned in warranty card).

WARRANTY

The blood-pressure monitor is guaranteed for 18 months
from date of purchase. This guarantee includes the
instrument and the cuff. The guarantee does not apply
to damage caused by improper handling, accidents, not
following the operating instructions or alterations made to
the instrument by third parties.
The guarantee is only valid upon presentation of the
guarantee card filled out by the representative of the sales
organization and its stamp and safety of the packaging.

TECHNICAL SPECIFICATIONS:

Weight:
Size:
Storage temperature:
Humidity:
12
537 g
200 õ 120 õ 75 mm
-5 î Ñ to +35 î Ñ
80% relative humidity
maximum
Operation temperature:
Measuring range:
Measuring resolution:
Accuracy:
Air leakage:
Accessories:
1. sphygmomanometer
2. cuff (adult with arm circumference of 25.4-40.6 cm)
with inlaid latex bladder
3. bulb and valve
4. stethoscope
5. soft bag
6. packaging
Technical alterations reserved!
0 tî 35 î Ñ
0 to 300 mmHg
2 mmHg
within ±3mmHg in 15 î Ñ to
25 î Ñ
within ±6mmHg in 34 î Ñ to
46 î Ñ
Sphygmomanometer
pointer in the zero zone is
the norm
<±3.75 mmHg/sec.
13