Major tech MT1887 User Manual - Page 15

Browse online or download pdf User Manual for Multimeter Major tech MT1887. Major tech MT1887 17 pages. True rms industrial multimeter

Major tech MT1887 User Manual
Function
Resistance
Capacitance
Frequency
(electronic)
Frequency
(electrical)
Duty Cycle
Temp
(type-K)
4-20mA%
Note: Accuracy specifications consist of two elements:
 (% reading) – This is the accuracy of the measurement circuit.
 (+ digits) – This is the accuracy of the analog to digital converter.
Range
Resolution
400
0.01
4k
0.0001k
40k
0.001k
400k
0.01k
4M
0.001M
40M
0.001M
40nF
0.001nF
400nF
0.01nF
4F
0.0001F
40F
0.001F
400F
0.01F
4000µF
0.1µF
40mF
0.001mF
40Hz
0.001Hz
400Hz
0.01Hz
4kHz
0.0001kHz
40kHz
0.001kHz
400kHz
0.01kHz
4MHz
0.0001MHz
40MHz
0.001MHz
100MHz
0.01MHz
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min
@ 20% to 80% duty cycle and > 100kHz.
40.00-400Hz
0.01Hz
Sensitivity: 15Vrms
0.1 to 99.90%
0.01%
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
-58 to 1832F
1F
-50
to 1000C
1C
-25 to 125%
0.01%
0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125%
Accuracy
(0.3% reading + 9 digits)
(0.3% reading + 4 digits)
(2.0% reading + 10 digits)
(3.5% reading + 40 digits)
(3.5% reading + 10 digits)
(5% reading + 10 digits)
(0.1% reading + 1 digits)
Not specified
(0.5% reading)
(1.2% reading + 2 digits)
(1.0% reading + 4.5°F)
(1.0% reading + 2.5°C)
(probe accuracy not included)
±50 digits