Centurion Scientific PrO-Hospital Ambient Instruction Manual - Page 7

Browse online or download pdf Instruction Manual for Laboratory Equipment Centurion Scientific PrO-Hospital Ambient. Centurion Scientific PrO-Hospital Ambient 20 pages.

Tube & bucket placements Swing out rotors
Small Swing out rotor
ALWAYS
load your tubes opposite to each other
ALWAYS
weigh samples to within 0.1 Gram. For best separations
ALWAYS
fit ALL the Tube Holders and any supports
CHECK The rotor screw fixing tightness regularly
.................................................................................................................................................................................
Medium – Large Swing out Rotors
ALWAYS
load your tubes opposite to each other
ALWAYS
Weigh Bucket and all samples within to 1.0 Gram
ALWAYS
fit ALL 4 Buckets and 4 Adaptors supplied
CHECK
the rotor screw fixing tightness regularly
.................................................................................................................................................................................
Microtitor Rotor
ALWAYS
load your plates opposite to each other
ALWAYS
weigh samples to within 1.0 Grams. For best separations
ALWAYS
fit both Buckets supplied (2)
CHECK the rotor screw fixing tightness regularly
7