Steps 6 through 8 are for the 3000AE only.
Connect the auxiliary dummy cell AEC1. Connect the five regular, colored leads to the top jacks, and the
multiple pairs of leads to the array of 16 jacks. Connect the other ends of the cables to the correct jacks
on the 3000AE.
Turn on the instrument and open Gamry
Framework™.
Select Experiment > Utilities > Aux
Electrometer Calibration
Select the radio button for the correct instrument, and click the OK button.
After you confirm several instructional
messages, the calibration proceeds
automatically, and you are notified if the
calibration is successful.
C
ALIBRATION
Did you receive a
Standard leads connect
in this row.
Auxiliary leads
connect from here
down.
Note: The multiple pairs of leads are
labeled with the channel numbers
and polarities. Be sure to match the
leads to the jacks.
W
ARNING?
TROUBLESHOOTING