1 .0 Safety Information
This instruction manual contains important operating and maintenance
instructions that must be read, understood, and followed by the product
user . Failure to use this product according to this instruction manual may
degrade or defeat the protection normally provided by this product . Read
this instruction manual prior to product use and keep this instruction
manual for future reference .
Product Symbols
Caution – Risk of Danger: Cautions there is material in the
instruction manual which must be read, understood, and
followed in order to preserve product safety features .
Caution – Hot Surface: Cautions that the top plate is too
hot to touch .
Indicates that the unit is plugged into power supply .
Indicates that the accessory external temperature
controller is properly plugged into the unit .
Warnings
Personal Injury
Do not use this product in a manner other than as stated in the Operating
w
Conditions section as the protection provided by the equipment may be
impaired .
This product is designed for use in laboratory environments by persons
w
knowledgeable in safe laboratory practices .
Always wear safety glasses and other appropriate protective equipment
w
when operating this product .
Electric Shock
This product must be connected to a grounded power outlet for safe
w
functioning .
Use only the power cord supplied with the product .
w
The power cord is the device available for full disconnect from mains input .
w
Position the product for use so the power cord can be easily disconnected
w
without having to move the product .
Disconnect the power cord before moving or cleaning the unit .
w
2
www.
GlobalScientificSupply
.com
Product Damage
Keep the product dry and clean .
w
Do not immerse the product for cleaning .
w
The ceramic top may break if impacted .
w
The maximum gross weight placed on the top surface must not exceed
w
25 lbs . (11 kg) .
These units are not explosion- or spark-proof .
w
Do not heat or stir volatile or flammable materials .
w
Do not operate this product near volatile or flammable materials .
w
Do not use this product with a metal vessel .
w
2 .0 Operating Conditions
Corning hot plates, stirrers, and stirrer/hot plates are designed to provide
safe functioning under the following conditions:
Indoor use
w
Altitude up to 6,500 feet (2,000 meters)
w
Ambient temperatures of 0°C to 30°C
w
Product should be placed on a flat surface at least 12 in . (30 .5 cm) from
w
walls, 48 in . (122 cm) from ceilings, and 12 in . (30 .5 cm) from other hot
plates if using multiple units .
Maximum relative humidity of 80% for temperatures up to 31°C, decreasing
w
linearly to 50% relative humidity at 40°C .
Pollution Degree 2: Any foreign matter that may accumulate on or within
w
the product during normal use is not electrically conductive .
Installation Category II: Product is designed for connection to an electrical
w
branch circuit inside a building with main supply voltage fluctuations not
exceeding ±10% of the nominal voltage .
GLOBAL SCIENTIFIC SUPPLY