3

Specifications

3.1

Technical data

Calibration use
Calibrated surface
Wavelength range
Calibration repeatability
Calibration relative uncertainty to NIST standard
Calibration valid for
Warm-up time
Bulb output
Power requirement
Dimensions
3.2

Ordering information

AvaLight-DH-CAL
AvaLight-DH-BAL-CAL
AvaLight-DH-CAL-ISP30
AvaLight-DH-CAL-ISP50
AvaLight-DH-CAL-ISP80
DH-Recal
AvaSoft-IRRAD
Revision: 1-00
© Avantes BV
Irradiance µW cm
CC-UV/VIS or AvaSphere
200-1099 nm
±1.0%
±10% (200 - 240 nm)
±9% (240 - 350 nm)
±10% (350 - 400 nm)
±9.5% (400 - 1100 nm)
60 hrs
Ca. 30 min
-2
80 µW cm
nm
5 µW cm
-2
nm
-1
100 – 240 VAC
315 x 165 x 140 mm
NIST traceable UV/VIS deuterium/halogen light
source with CC-VIS/NIR diffuser
As AvaLight-DH-CAL, but balanced UV/VIS
deuterium/halogen light
NIST traceable UV/VIS deuterium/halogen light
source for use with AvaSphere-30-IRRAD, incl.
special sphere holder bottom plate
As AvaLight-DH-CAL-ISP30, but for use with
AvaSphere-50-IRRAD
As AvaLight-DH-CAL-ISP30, but for use with
AvaSphere-80-IRRAD
AvaLight-DH-CAL reclibration service (200 – 1099
nm)
Irradiance application add-on software, to be
ordered with AvaSoft-Full
www.avantes.com
-2
-1
nm
-1
(@ 215 nm)
(@ 800 nm)
6