8. Turn the key to the in the back of the unit to the on position, turn the dial on the front of the control
box to the desired laser power (Note: the readout on the unit is in current, not power) and press the
laser enable button to turn on the laser inside of the probe.
Note that this laser is configured with a remote interlock. IPS has provided an RJ-11 (4-pin telephone jack)
that is pre-wired to provide a closed loop connection for the interlock. Removal of the RJ-11 connector or
break in the closed loop circuit provided will result in a laser interrupt and the system will be reset to
standby mode. Users may reconfigure the RJ-11 connector to provide a closed loop that interactively couples
with their system or facility by replacing the short length of wire with longer closed loop connections if
desired.
01-04-16
Avantes
AvaRaman Bundle Manual
www.avantes.com
Enable Laser
RJ-11
Interlock
8
and collect
Spectra