4.
Coefficient Calibration
On the LED screen the internal counts are shown as 1/10 with scaling 0 to 430,000 counts for 2mV/V
The coefficient is same as screen value and 0.1 is cause to be a calibration to 43.000 screen value for 2mV/V
0mV/V --> 0 counts
2mV/V--> 430.000 counts (As LED screen)
The examples below are given for 2mV/V input
Coefficient
Screen
1.00000
430000
0.10000
43000
0.01000
4300
0.02326
10000
0.00698
3000
5.
Error Messages-Codes
Error No
Description
Hata No
ERR.005
"LC.HATA"
ERR.090
"ADCHat "
ERR.001
"MAK.YUK"
ERR.099
"DEG.COK"
L60 & L60Tr Technical Manual -
Katsayı Kalibrasyonu
Hata Mesajları-Kodları
Meaning
Tanım
Anlamı
Loadcell is overloaded or not installed. Check the connection and the
load value
ADC unit of the device is broken. Needs service
The screen value is above the Maximum Capacity
Parameteter Error, entered value is above the limits
L60 & L60Tr Teknik Manuel