a&p instruments UX 3081WQ Operating Instructions Manual - Page 16

Browse online or download pdf Operating Instructions Manual for Measuring Instruments a&p instruments UX 3081WQ. a&p instruments UX 3081WQ 20 pages. Moisture analysers

Operating instructions for moisture analysers
UX 3081
UX 3081WQ
Potential errors
Error
Display
Display blank:
time
mode
value (weight)
unit
time
mode
value (weight)
unit
Scale continuing to run:
Menu
Drying parameters cannot be
entered, only
temp. C°
time
unit
mode
value (weight)
Menu lock cannot be
unlocked
Scale
Scale cannot be tared
Scale cannot be calibrated:
temp. C°
time
mode
value (weight)
unit
a&p instruments, Kluckhofer Weg 61, D-32657 Lemgo, Tel.:05232 97780, Fax.: 977820
E-Mail: [email protected], Internet: www.apinstruments.de
Reason
Initial load too small or large,
Tray support and / or tray missing
Location with too much disturbance
Draught
Area around tray support is dirty
Scale is faulty
Scale without power
Scale is not connected to control card
Scale is faulty
Location with too much disturbance
Draught
Area around tray holder is dirty
Scale is faulty
Menu is locked
5-digit pin for unlocking menu lock is
incorrect
Weight within minus range
Weight difference less than 16 mg
or more than 16 mg
Initial load too small or large, tray holder
and / or tray missing.
Location with too much disturbance
Draught
Area around tray holder is dirty
Scale setting is incorrect
Scale is faulty
System error
Calibration process has stopped
Solution
Place empty drying dray on centre of
tray support,
Place on a vibration-free surface
Protect from draughts
Carefully clean area around tray support
Contact a&p instruments
Contact a&p instruments
Contact a&p instruments
Contact a&p instruments
Place on a vibration-free surface
Protect from draughts
Carefully clean area around tray holder
Contact a&p instruments
Unlock menu lock,
contact the person responsible or
a&p instruments
Contact a&p instruments at
Switch off device, place empty tray on
balance and switch on again
Less than 16 mg set to zero, press
button briefly
More than 16 mg tare, press and hold
Place empty drying dray on centre of
tray holder,
Place on a vibration-free surface
Protect from draughts
Carefully clean area around tray holder
Contact a&p instruments
Contact a&p instruments
Disconnect device from mains and switch
on again.
Seite 16