A-M Systems 2100 Instruction Manual - Page 14

Browse online or download pdf Instruction Manual for Measuring Instruments A-M Systems 2100. A-M Systems 2100 19 pages. Isolated pulse stimulator

A-M Systems 2100 Instruction Manual
There are three classes of specifications. Class A specifications tested in all units, and
are guaranteed. Class B specifications are inherent in the design; they are indirectly
tested and are guaranteed. Class C specifications are typical operating values which
are occasionally tested-, these are typical values, given for your information but not
guaranteed. The class for each specification is noted in the center column of the
following sections.
Inputs
Switching level
Hysteresis
Input equivalent circuit
Minimum duration
(TTL-level signal)
Open circuit condition
Timing and rate accuracy
Rate selector steps
Timing multiplier steps
Resolution
Timing error (second and later
pulses after resetting controls)
Timing error (first pulse after
readjusting timing controls)
Timing repeatability
A-M Systems

Specifications

A
C
C
C
B
B
B
B
B
C
C
131 Business Park Loop, P.O. Box 850 Carlsborg, WA 98324
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website:
1.6 V (± 0.2 V)
approx. 50 mV
100KΩ to +5 in parallel
with no more than 35pF
> 50 ns.
defaults to high/on level
100µs to 100s in 7 steps
1% of rate selector step
size. Maximum value is
9.99 x the rate selector
step size
approx. 0.1% of the maximum
value
< 0.02% of rate selector value +
0.15 µs
an additional 0.25 µs error may
occur
< 0.01% setting + 20 ns
http://www.a-msystems.com
11