Chapter 5: Calibration
5.1 Calibrating the Orion
Press the Dose button rapidly 5 times in the Weighing Scale Mode to enter the calibration
mode.
Place the 500-gram calibration weight on the platform and allow the Orion to calibrate.
1. While the Orion is in Weighing Scale Mode and the weighing unit is in grams, rapidly
press the Dose button multiple times until the LED display shows [ CAL ] and then
[ 500 ].
2. Place a 500-gram calibration weight on the platform and wait for the Doser to execute
the calibration procedure.
3. The display shows [ -End- ] when the calibration procedure is completed.
4. You may unload the calibration weight when completed.
* Note: Calibration weights of 100 g, 1000 g, 1500 g, 2000 g, and 3000 g can also be
used. For highest accuracy use Acaia calibration weights or F1 or F2 Class calibration
weights. Inexpensive lower-class calibration weights have a higher deviation from a true
accurate weight.