Oxygen Cylinder Depletion Times
These times are approximate and assume full cylinder capacity. Always monitor the
cylinder pressure and low pressure alarm to make sure you do not run out of oxygen.
E Cylinder
8
9
1200
67
60
1000
80
72
800
98
88
600
127
115
500
149
135
400
180
163
300
225
206
200
293
274
Tidal
Volume
Jumbo D Cylinder
8
9
1200
63
56
1000
75
67
800
92
82
600
119
107
500
139
126
400
168
152
300
210
192
200
274
256
Tidal
Volume
Capacity =
682
Breaths Per Minute
10
12
14
54
46
39
65
54
47
80
67
58
104
88
76
123
104
90
149
127
111
189
163
143
256
225
200
Capacity =
637
Breaths Per Minute
10
12
14
51
43
37
61
51
44
75
63
54
97
82
71
115
97
85
139
119
104
177
152
134
239
210
187
Liters Oxygen Capacity
(4.6 Liters Water Capacity)
15
18
20
37
44
54
72
60
54
85
72
65
104
88
80
135
115
104
189
163
149
Liters Oxygen Capacity
(4.0 Liters Water Capacity)
15
18
20
34
41
51
67
56
51
79
67
61
97
82
75
126
107
97
177
152
139
14
22
24
26
50
46
42
59
54
50
73
67
62
96
88
82
137
127
119
22
24
26
46
43
40
55
51
47
68
63
58
89
82
76
128
119
111
28
30
39
37
47
44
58
54
76
72
111
104
28
30
37
34
44
41
54
51
71
67
104