CPS IAQPRO SmartAir Quick Start And Troubleshooting Manual - Page 3

Browse online or download pdf Quick Start And Troubleshooting Manual for Measuring Instruments CPS IAQPRO SmartAir. CPS IAQPRO SmartAir 8 pages. Professional indoor air quality meter
Also for CPS IAQPRO SmartAir: Owner's Manual (18 pages), Owner's Manual (9 pages)

CPS IAQPRO SmartAir Quick Start And Troubleshooting Manual

QUICK START INSTRUCTIONS

Step 2: (5 Minutes Required) Relocate the unit to a secure, dry
location OUTDOORS with access to power (do not locate in
direct sunlight). It is OK to disconnect the unit from the INDOOR
power source while relocating it to the OUTDOOR location.
Power unit ON and wait for 5 mins.
Step 3: Open the CPS Link App on your smart device and select the
IAQPRO icon to connect unit. Once connected, online guidance
screens will appear prompting completion of Step 1 & 2 above.
Swipe right to advance to the SET BASELINE screen. Touch
the SET BASELINE option to save tVOC baseline and complete
setup.
CONDUCT AN INDOOR AIR QUALITY TEST
The tVOC sensor requires a stabilization period when powered ON (the light bar will flash
Green/Yellow during this process). The stabilization period can last between 5 to 60 minutes
depending on the combination of storage conditions and time (minutes, days, weeks) the
unit was powered OFF. Unit can be powered ON during travel between jobs to minimize
jobsite stabilization time using available automobile USB power connections or other readily
available portable power devices.
a. For more accurate readings and to generate an Indoor Air Quality Test Report, start a
30-minute test (default setting).
• Choose the "TESTS" icon (upper right of screen), then.
• Choose the "Air Quality" option (lower left of screen).
• Choose a "Standard Test" (30 minutes) or "Custom Test".
• If "Standard Test" selected, touch "Start Test" button to begin. A new screen will
appear, and the timer counts backwards from 30 minutes.
• To conduct a test shorter or longer than 30 minutes, select "Custom Test." Enter
the desired test period (minimum of 5 min to a maximum of 23 hrs, 59 mins, 59
seconds), then press "Start Test".
b. The light bar on the SmartAir housing will display a
color indicating the range (see chart at right) within
which the highest reading for ONE of these values
(PM
, PM
2.5
10
SmarAir is paired to your mobile device with the CPS
Link app, it will emit a a short blue flash once every
5 seconds.
c. For more detailed information about any item being
measured, see the Information Section in the app.
d. Users are encouraged to pair the SmartAir to a mobile device using the CPS Link app
for more detailed information about the quality of indoor air.
, tVOC or CO
2
) is found. Note: If the
GREEN
Good
YELLOW
Satisfactory
DARK YELLOW
Moderate
ORANGE
Poor
RED
Severe
+
3