Miele PG 8583 CD Installations Plan - Page 5

Browse online or download pdf Installations Plan for Washer Miele PG 8583 CD. Miele PG 8583 CD 14 pages. Laboratory glassware washer

Miele PG 8583 CD Installations Plan
Electrical connection (3 x 20 A)
Voltage (default)
Rated load
Fuse rating
Minimum cross-section
Length of electrical cord (UL 921) with plug
Electrical connection
Voltage (convertible)
Rated load
Fuse rating
Minimum cross-section
Length of electrical cord (UL 921) with plug
Electrical connection
Cold water connections* (KW+DK)
Max. temperature
Max. water hardness
Recommended flow pressure
Min. flow pressure
Max. pressure
Throughput
Connection thread - on site (flat sealing)
Length of connection hose (supplied)
Length of the steam condensor hose
* Two cold water connections required: use supplied Y-adapter if only one cold water connection is available.
Hot water connection (WW)
Max. temperature
Max. water hardness
Recommended flow pressure
Min. flow pressure
Max. pressure
Throughput
Connection thread - on site (flat sealing)
Length of connection hose (supplied)
 
Installation plan PG 8582 CD / PG 8583 CD 
 
 
3 AC 208V 60 Hz
6.2 kW
3 × 20 A
4 × AWG 12
approx. 5'9"
NEMA L15-20
2 AC 208V 60 Hz
6.2 kW
2 × 30 A
3 × AWG 10
approx. 5'9"
NEMA L6-30
68 °F
73 gpg
29 psi
14.5 psi
145 psi
2 gal/min
3/4" Male Garden Hose
Thread
(USA: 11.5 NH)
approx. 5' 7"
approx. 5' 7"
150 °F
73 gpg
29 psi
5.8 psi
145 psi
2 gal/min
3/4" Male Garden Hose
Thread
(USA: 11.5 NH)
approx. 5' 7"
3 AC 208V 60 Hz
6.2 kW
3 × 20 A
4 × AWG 12
approx. 1.8 m
NEMA L15-20
2 AC 208V 60 Hz
6.2 kW
2 × 30 A
3 × AWG 10
approx. 1.8 m
NEMA L6-30
20 °C
70 °d
200 kPa
100 kPa
1,000 kPa
7.5 l/min
3/4" Male Garden
Hose Thread
approx. 1.7 m
approx. 1.7 m
65 °C
70 °dH
200 kPa
40 kPa
1,000 kPa
7.5 l/min
3/4" Male Garden
Hose Thread
approx. 1.7 m
page 4 / 14