3B Medical Lumin LM3000 User Manual - Page 5
Browse online or download pdf User Manual for Personal Care Products 3B Medical Lumin LM3000. 3B Medical Lumin LM3000 7 pages. Instructions for healthcare facilities: bioburden reduction uv system user's manual
Also for 3B Medical Lumin LM3000: Instructions Manual (14 pages)
ELECTROMAGNETIC COMPATIBILITY
Item
Compliance
Radiated Emission
P
(CISPR 11)
Harmonics of Current
P
(IEC 61000-3-2.)
Voltage Flicker
–
(IEC 61000-3-3)
Electrostatic
Discharge Immunity
P
(IEC 61000-4-2)
Conducted
Disturbance Immunity
P
(IEC 61000-4-6)
Electrical Fast
Transient/BurstIEC
P
(61000-4-4)
Surge Immunity
P
(IEC 61000-4-5)
Voltage Dips and
Interruptions
P
(EN 61000-4-11)
Radiated
Electromagnetic Field
P
Immunity
(IEC 61000-4-3)
Power Frequency
Magnetic Field
P
Immunity
(IEC 61000-4-8)
8
Electromagnetic
Environment
Its RF emissions are very
low and are not likely to
cause any interference
in nearby electronic
equipment
40 order harmonics
Without testing according
to clause 6.1 of IEC
61000-3-3.
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15
kV air
Home Healthcare
Environment
±2 kV 100 kHz repetition
frequency
Line to line±0.5kV, ±1Kv
Line to earth ±0.5kV,
±1kV,±2kV
0%UT 0.5(at 0°, 45°, 90°,
135°, 180°, 225°, 270°,
315°)
0%UT 1(at 0°)
70%UT 25(at 0°)
0%UT 250(at any angle)
Home Healthcare
environment
80 to 2700 MHz (Front,
Rear, left, right)
Field Strength: 10V/M
Orientations of Induction
Coil (X, Y, Z)
Magnetic Field Strength:
30A/m
Warning - the Lumin is not a toy and is not intended to be
used by children.
Warning - the Lumin has a powerful UV light bulb
included, use with care in case of broken glass.
Warning - do not use in or around sinks, tubs, pools or any
bodies of water.
Warning - the UV that is emitted in the closed chamber is
not compatible with life, never place living creatures in the
Lumin.
Warning - Ultraviolet (UV-C) Radiation Hazard. Any
exposure to the UV-C will cause significant eye damage
and may cause skin damage. If you see any crack or
damage in the tinted oval window on the top of the unit,
unplug the power and stop using immediately, and contact
3B Medical for assistance.
Warning - do not stack objects on top of your Lumin.
Warning - use this device only for its intended use as
described in this manual. If the equipment is not used
in the manner specified, the cleaning provided by the
equipment may be less effective.
Warning - Do not attempt to operate this device with
the drawer open. The device was designed to shut off
automatically when the drawer is open. It is important not
to defeat or tamper with this safety feature as UV light can
cause damage to eyesight.
Warning - The Lumin plugs into an outlet. Disconnect the
device by unplugging. Do not plug the device into an outlet
that is difficult to access.
Warning - Lumin relies on the optics of UVC light. Results
may vary depending on positioning and shadowing.
9