All-layer sample
1. Slide the wiping ring to the desired sampling
depth.
2. Push the push button and hold it. Submerge the
instrument into the medium to be sampled up to the
desired depth. Thereby, the instrument is filled with
the liquid to be sampled.
3. Release the push button. Thereby, the bottom valve
closes.
4. Remove the instrument from the medium to be
sampled.
5. Empty the sample into a suitable collecting cont-
ainer by pushing the push button.
After taking the sample
1. Slide the wiping ring with the aid of the supplied
pusher down to wipe any remaining liquids.
2. Screw off the discharge funnel .
3. Hold the valve cone and screw off the push button.
4. Pull all parts out of the sampling tube.
5. Clean all parts. Use the cleaning brushes (see acces-
sories) for the inside of tubes.
Troubleshooting
Problem
The instrument doesn't
fit through the container
opening.
Bottom valve is dripping.
Accessories
Type
Cleaning brush
Drain safety stand
Sample bottle HDPE
Sample bottle HDPE
More accessories can be found at www.buerkle.de
To eliminate the error
Srew off the discharge funnel and take the sample without it. Screw the discharge
funnel on again before emptying the sample.
1. Bottom valve is soiled. Clean the instrument.
2. Discharge funnel is screwed on too tightly and the valve seat is deformed.
Umscrew the discharge funnel.
Specification
PVC/stainless steel, 100 cm
Total length 117 cm, Ø 35 mm
250 ml
500 ml