4. Remove the instrument from the medium to be
sampled.
5. Empty the sample into a suitable collecting cont-
ainer by sliding in the piston rod.
After taking the sample
1. Slide the wiping ring with the aid of the supplied
pusher down to wipe any remaining liquids.
2. Screw off the discharge funnel .
3. Screw off the guide nut .
4. Pull the handle and pull out the piston rod comple-
tely.
5. For cleaning purposes, the hexagon screw of the
suction piston can be opened by using the supplied
key.
6. Clean all parts. Use the cleaning brushes (see acces-
sories) for the inside of tubes.

Troubleshooting

Problem
The instrument doesn't
fit through the container
opening.
Poor suction power or
suction piston is leaky.
Accessories
Type
Cleaning brush
Drain safety stand
Sample bottle HDPE
Sample bottle HDPE
More accessories can be found at www.buerkle.de
To eliminate the error
Srew off the discharge funnel and take the sample without it. Screw the discharge
funnel on again before emptying the sample.
Increase the contact pressure of the suction piston. Screw off the discharge funnel
and adjust the o-rings clockwise with a hexagon key (the o-rings must seal tightly
and should have a low friction).
Specification
PVC/stainless steel, 100 cm
Total length 117 cm, Ø 35 mm
250 ml
750 ml