DX-300 – DIVER HELD MAGNETOMETER
Symptom:
Erratic Performance from DX-300 Module
1.
First check as above.
2.
Check that the contacts of the bulkhead connectors are clean.
3.
Check that the connectors of the Sensor Probe and Bonephone are
clean. If necessary reduce the contact resistance by slightly prising the
gap in the male contacts, this will increase the firmness of the
connector on insertion.
4.
Check for visible damage to Sensor Probe cable and connector, which
could have caused water ingress.
5.
Check that the male connector seals are in place and clean.
6.
If you are unable to determine the cause of the erratic behaviour,
contact your Aquascan agent for assistance.