Dickson MM125 User Manual - Page 2

Browse online or download pdf User Manual for Thermometer Dickson MM125. Dickson MM125 2 pages. Alarm thermometer

Clearing Alarm Conditions
1.
When the unit enters alarm mode:
a. The Audio Alarm will sound for 30
seconds
b. The alarm symbol will flash, and the out of
range high or low reading will flash.
c. If no action is taken to clear the alarm, the
audio alarm will beep 2x every 5 minutes and
the display will continue to flash until the clear
button is pressed. This condition will continue
even if the temperature returns to normal.
2.
Clearing an Alarm
a. Press the clear button to temporarily stop the
audible alarm. The display will continue
flashing. If the alarm condition still exist, the
unit will sound again once the sample rate
period has passed.
b. Once the unit moves out of the alarm condition,
press and hold the Clear button until CLR is
displayed. This will reset the alarm and stop
the flashing on the display.
www.DicksonData.com • 930 South Westwood Avenue • Addison, IL 60101-4917
Telephone: 630.543.3747 • Fax: 630.543.0498 • E-mail: [email protected]
3.
Alarm Notification
a. If the alarm is not on (dip switch is down), the
display will flash if the temperature moves into
alarm range.
Note:
If no alarm notification is desired, change alarm
settings to high/low outer range limits (Low=
-58 deg F, High= 158 deg F)
b. Once the temperature moves out of alarm
range, press and hold the clear button until CLR
is displayed to reset
Rev. 07/08