a. the pulses are time continuant
b. pulse/ pause 1:1
c. maximal pulse-time are programmable
Test pulse rate:
q
p
0,6
1
1,5
2,5
3,5
6
10
15
25
40
60
100
Volume test and calibration by NOWA
2.5.
The meter test and calibration over NOWA is available.
3. Energy test and calibration
Manual energy test
3.1.
For the energy test it is not necessary to set the meter into calibration mode.
a) Setting into energy test via the button on the meter
1. Heat up temperature sensor or connect measuring shunt.
2. Change to the power display (with button; do not keep button pressed).
123.456
kW
3. Press button for approx. 10 s until "Et1" display appears (ignore the loop change!).
Et 1
Industriestraße 13
91522 Ansbach
Deutschland
Tel. +49 981 18 06-015
www.diehl.com/metering
Pulse rates
1ml
167Hz @qp
1ml
278Hz @qp
2ml
208Hz @qp
3ml
231Hz @qp
4ml
243Hz @qp
6ml
278Hz @qp
10ml
278Hz @qp
20ml
208Hz @qp
30ml
231Hz @qp
40ml
278Hz @qp
60ml
278Hz @qp
60ml
463Hz@qp