Keypad
The BondMaster 600 is available with an English, International, Chinese, or Japanese keypad
configuration.
English keypad
Chinese keypad
English
Enter
Return
CAL/NULL
GAIN
BondMaster 600 — Getting Started Guide
International
The Enter key is used to make selections.
The Return (or, Back) key is used to exit a menu and
return to a previous screen.
If this direct function key is pressed once, it nulls
(zeros) the BondMaster 600. If this key is pressed and
held, a calibration wizard is initiated in RESON
(resonance) and MIA (mechanical impedance
analysis) modes of operation only.
Direct function key used to display either the
BondMaster 600 combined horizontal and vertical
gain settings, horizontal gain only, or vertical gain
only.
International keypad
Japanese keypad
Function