Prior Lumen200 Metal Halide Lamp Operation:

Starting Up the Lumen
1. Switch the Lumen power switch on.
2. Make sure the ventilation vent on the left hand side is unobstructed or the lamp
will overheat resulting in automatic shutdown and damage to the module.
3. Allow 1-5 minutes for light to reach 70% of output.
4. Allow 30 minutes for the Lumen to reach operational temperature.
5. Warning: Do not power down the unit within 30 mins of power up. This may re-
duce the effective lifetime of the bulb.
Shutting down the Lumen
The following warnings apply as damage to the bulb may result if instructions not
followed:
1. Warning: Do not shut the unit down within 30 minutes of powering up the unit.
2. Warning: After shutting down the unit allow 30 minutes before re-powering up or
changing the bulb. Failure to do so is likely to result in damage to the bulb.
Warning: the air
outlet for heat
ventilation must
not but obstructed